Diagnostic accuracy of flat-panel computed tomography in assessing cerebral perfusion in comparison with perfusion computed tomography and perfusion magnetic resonance: a systematic review
Publication year
2019Source
Neuroradiology, 61, 12, (2019), pp. 1457-1468ISSN
Publication type
Article / Letter to editor

Display more detailsDisplay less details
Organization
Medical Imaging
Neurosurgery
Radboudumc Extern
Health Evidence
Operating Rooms
Journal title
Neuroradiology
Volume
vol. 61
Issue
iss. 12
Page start
p. 1457
Page end
p. 1468
Subject
Radboudumc 10: Reconstructive and regenerative medicine RIHS: Radboud Institute for Health Sciences; Radboudumc 15: Urological cancers RIHS: Radboud Institute for Health Sciences; Radboudumc 9: Rare cancers RIHS: Radboud Institute for Health Sciences; Health Evidence - Radboud University Medical Center; Medical Imaging - Radboud University Medical Center; Neurosurgery - Radboud University Medical Center; Operating Rooms - Radboud University Medical CenterAbstract
PURPOSE: Flat-panel computed tomography (FP-CT) is increasingly available in angiographic rooms and hybrid OR's. Considering its easy access, cerebral imaging using FP-CT is an appealing modality for intra-procedural applications. The purpose of this systematic review is to assess the diagnostic accuracy of FP-CT compared with perfusion computed tomography (CTP) and perfusion magnetic resonance (MRP) in cerebral perfusion imaging. METHODS: We performed a systematic literature search in the Cochrane Library, MEDLINE, Embase, and Web of Science up to June 2019 for studies directly comparing FP-CT with either CTP or MRP in vivo. Methodological quality was assessed using the QUADAS-2 tool. Data on diagnostic accuracy was extracted and pooled if possible. RESULTS: We found 11 studies comparing FP-CT with CTP and 5 studies comparing FP-CT with MRP. Most articles were pilot or feasibility studies, focusing on scanning and contrast protocols. All patients studied showed signs of cerebrovascular disease. Half of the studies were animal trials. Quality assessment showed unclear to high risks of bias and low concerns regarding applicability. Five studies reported on diagnostic accuracy; FP-CT shows good sensitivity (range 0.84-1.00) and moderate specificity (range 0.63-0.88) in detecting cerebral blood volume (CBV) lesions. CONCLUSIONS: Even though FP-CT provides similar CBV values and reconstructed blood volume maps as CTP in cerebrovascular disease, additional studies are required in order to reliably compare its diagnostic accuracy with cerebral perfusion imaging.